Channel Name | Low Filter (Hz) | High Filter (Hz) | Display Polarity | Display Amplitude |
---|---|---|---|---|
E1-REF | 0.3 | 35 | Negative ↑ | * |
E2-REF | 0.3 | 35 | Negative ↑ | * |
Lchin-REF | 10 | 70 | Negative ↑ | * |
Rchin-REF | 0.3 | 35 | Negative ↑ | * |
Cchin-REF | 0.3 | 35 | Negative ↑ | * |
C3-REF | 0.3 | 35 | Negative ↑ | * |
C4-REF | 0.3 | 35 | Negative ↑ | * |
O1-REF | 0.3 | 35 | Negative ↑ | * |
O2-REF | 0.3 | 35 | Negative ↑ | * |
F3-REF | 0.3 | 35 | Negative ↑ | * |
F4-REF | 0.3 | 35 | Negative ↑ | * |
T3-REF | 0.3 | 35 | Negative ↑ | * |
T4-REF | 0.3 | 35 | Negative ↑ | * |
M1-REF | 0.3 | 35 | Negative ↑ | * |
M2-REF | 0.3 | 35 | Negative ↑ | * |
ECG1-REF | 1.0 | 35 | Positive ↑ | * |
ECG2-REF | 1.0 | 35 | Positive ↑ | * |
ECG3-REF | 1.0 | 35 | Positive ↑ | * |
LLeg1-REF | 10 | 70 | Negative ↑ | * |
LLeg2-REF | 10 | 70 | Negative ↑ | * |
RLeg1-REF | 10 | 70 | Negative ↑ | * |
RLeg2-REF | 10 | 70 | Negative ↑ | * |
Airflow | 1.0 | 15 | Positive ↑ | * |
CannulaFlow | 0.05 | 15 | Positive ↑ | * |
SUM | 0.05 | 15 | Positive ↑ | * |
Chest | 0.05 | 15 | Positive ↑ | * |
ABD | 0.05 | 15 | Positive ↑ | * |
* Set Display Amplitude to adequately visualize all waveforms
Signal Type | Signal Inputs | Channel Name | Low Filter (Hz) | High Filter (Hz) | Display Polarity | Display Amplitude |
---|---|---|---|---|---|---|
EOG | E1-REF | E1 | 0.3 | 35 | Negative ↑ | * |
EOG | E2-REF | E2 | 0.3 | 35 | Negative ↑ | * |
Chin EMG | Lchin-Cchin | Chin | 10 | 70 | Negative ↑ | * |
EEG | F4-M1 | F4-M1 | 0.3 | 35 | Negative ↑ | * |
EEG | C4-M1 | C4-M1 | 0.3 | 35 | Negative ↑ | * |
EEG | O2-M1 | O2-M1 | 0.3 | 35 | Negative ↑ | * |
EEG | T4-M1 | T4-M1 | 0.3 | 35 | Negative ↑ | * |
EEG | F3-M2 | F3-M2 | 0.3 | 35 | Negative ↑ | * |
EEG | C3-M2 | C3-M2 | 0.3 | 35 | Negative ↑ | * |
EEG | O1-M2 | O1-M2 | 0.3 | 35 | Negative ↑ | * |
EEG | T3-M2 | T3-M2 | 0.3 | 35 | Negative ↑ | * |
ECG | ECG3-ECG1 | ECG | 1.0 | 35 | Positive ↑ | * |
Oximetry-Pulse | Internal DC | Pulse | * | |||
Oximetry Plethysmography Waveform or other quality indicator | Pleth (from Oximetry) or other quality indicator | Pleth or name of quality indicator | * | |||
Limb EMG | LLeg1-LLeg2 | LLeg | 10 | 70 | Negative ↑ | * |
Limb EMG | RLeg1-RLeg2 | RLeg | 10 | 70 | Negative ↑ | * |
Snore Mic (if used) | Bipolar | Snore | 20 | 100 | Positive ↑ | * |
Nasal/Oral Flow | Bipolar | Airflow | 1.0 | 15 | Positive ↑ | * |
Nasal Pressure | Pressure Transducer | CannulaFlow | 0.05 | 15 | Positive ↑ | * |
IP SUM | Bipolar | SUM | 0.05 | 15 | Positive ↑ | * |
IP Chest Effort | Bipolar | Chest | 0.05 | 15 | Positive ↑ | * |
IP Abdominal Effort | Bipolar | ABD | 0.05 | 15 | Positive ↑ | * |
Oximetry numerical | Internal DC | SAO2 | * | |||
capnography | Auxiliary DC | Cap | * | |||
capnography | Auxiliary DC | EtCO2 | * | |||
Body Position | Internal DC | Position | * |
* Set Display Amplitude to adequately visualize all signals